Focus On Flow
 
Our Free Newsletter 
Sign Up Now!

 

dot dot Model 268 Full-Pipe Flo-Tote

Home
Multi-Mag
Flo-System
Model 268

 
Measurement
Volumetric flow in full pipes, 2" (5 cm) to 96" (244 cm.) in diameter using insertable velocity sensor. Flow indication in English or Metric units.

Local Velocity Measurement
Method: Electromagnetic (Faraday's Law)
Range: -5 to +20 ft/s (-1.5 to +6.1 m/s)
Zero Stability: ±0.05 ft/s (±15.2 mm/s)
Accuracy: ±2% of reading ±zero stability at -3 to +10 ft/s (-0.9 to +3 m/s)
Resolution: 0.01 ft/s (3.05 mm/s)

Sensor Selection
Pipe Sizes 2" (5 cm) to 30" (76 cm) - 1" sensor (Recommended for clean water applications only up to 30". Clean water applications over 30" should use 2" sensor)
Pipe Sizes 6" (15 cm) to 96" (244 cm) - 2" sensor

Materials
Sensor: Polyurethane exposed to flow Epoxy ceramic sensors available for chemical applications. 20 ft., polyurethane outer jacket. Additional cable lengths available - Contact factory.

Sensor Mounting Hardware
1" Sensor: 316 Stainless Steel
2" Sensor: Carbon Steel and PVC exposed to fluid; Stainless Steel Optional

Environment - Operating Temperature Degree Range
Electronics: 32°F to 113°F (0°C to 45°C)
Sensor: 32°F to 160°F (0°C to 71°C) @ 250 PSI with Stainless Steel Hardware
Consult factory for information on higher temperature ranges or sensors for extreme environments.

Data Retrieval
Portable computer utilizing MS/DOS is utilized for Flo-Ware (sold separately) operating software for Flo-Tote flowmeter.

Power Requirements
Six Volt (primary or rechargeable) battery
Power Consumption: 110 mA @ 6 volts.

Pressure/Temperature Limits
PVC Insertion Tube: 150 PSI @ 105°F (41°C),
30 inch max. (56 cm)
Stainless Steel Insertion Tube: 250 PSI @ 160°F (71°C), 65 inch max. (165 cm)

Pressure Transducer (Optional)
Range: (Choice of transducers) 0-250 psi or 0-100 psi
Accuracy: ±3% of reading, or ±1 PSI, whichever is greater.
Housing
Sealed, watertight PVC enclosure (submersible)
Length: 21.5"; Diameter 5"
Weight: 15 lbs. with battery

Flow Proportional Contact Closure (Qty. 2)
Contact closes each time flow total exceeds a preset value. (Limit 1 closure per cycle)
Minimum Value: 1 Gallon
Maximum Value: 99,000,000 Gallons
Contact Closure: 1 amp @ 28 volts, non-inductive
Closure Period: 100-300 milliseconds

Extra Data Logging Channels
Voltage Input Channels - (Qty. 4)
Voltage Range: -5 volts to +5 volts
Voltage Resolution: 2.5 mV

Pulse Input Channel - (Qty. 1)
Requires 3-15 VDC external pulse at maximum rate not to exceed 4095 pulses per data cycle.

External Start
+5 to +15 VDC, 1 second

Memory Size

Standard        (21000 cycles)
                      (# active channels)
i.e., if two voltage channels are recorded, velocity and level or velocity and pressure, number of active channels = 2. 21000 ÷ 2 = 10500 cycles.

Optional - 86,520 cycles

CONTACT FACTORY FOR ORDERING INFORMATION

 

dot dot

Model 268 Full-Pipe Flo-Tote

dot

 
Address: Marsh-McBirney, Inc. 4539 Metropolitan Ct. Frederick, Maryland 21704 USA
Telephone:
U.S. and Canada 1-800-368-2723 - Outside the U.S. 1-301-874-5599 - Fax: 1-301-874-2172

Copyright Marsh-McBirney, Inc.  All rights reserved.